Apogee Instruments PYRANOMETER SP-421 Руководство пользователя - Страница 7

Просмотреть онлайн или скачать pdf Руководство пользователя для Измерительные приборы Apogee Instruments PYRANOMETER SP-421. Apogee Instruments PYRANOMETER SP-421 18 страниц.

Apogee Instruments PYRANOMETER SP-421 Руководство пользователя
Spectral Response
Temperature Response
Spectral response estimate of Apogee silicon-cell
pyranometers. Spectral response was estimated
by multiplying the spectral response of the
photodiode, diffuser, and adhesive. Spectral
response measurements of diffuser and adhesive
were made with a spectrometer, and spectral
response data for the photodiode were obtained
from the manufacturer.
Mean temperature response of four Apogee
silicon-cell pyranometers. Temperature response
measurements were made at approximately 10 C
intervals across a temperature range of
approximately -10 to 50 C under sunlight. Each
pyranometer had an internal thermistor to
measure temperature. At each temperature set
point, a reference blackbody pyranometer was
used to measure solar intensity.
7