DEPLOYMENT AND INSTALLATION

Mount the sensor to a solid surface with the nylon mounting screw provided to prevent galvanic corrosion. To
accurately measure illuminance incident on a horizontal surface, the sensor must be level. An Apogee Instruments
model AL-100 Leveling Plate is recommended to level the sensor when used on a flat surface or being mounted to
surfaces such as wood. To facilitate mounting on a mast or pipe, the Apogee Instruments model AL-120 Solar
Mounting Bracket with Leveling Plate is recommended.
Nylon Screw: 10-32x3/8
Nylon Screw: 10-32x3/8
Model AL-120
Model AL-100
To minimize azimuth error, the sensor should be mounted with the cable pointing toward true north in the
northern hemisphere or true south in the southern hemisphere. Azimuth error is typically less than 1 %, but it is
easy to minimize by proper cable orientation.
In addition to orienting the cable to point toward the nearest pole, the sensor should also be mounted such that
obstructions (e.g., weather station tripod/tower or other instrumentation) do not shade the sensor. Once
mounted, the green cap should be removed from the sensor. The green cap can be used as a protective covering
for the sensor when it is not in use.