5.1.2 Dip Systems

The procedure is identical to that for the Floating Ball System
except that, in this case, it is only necessary to remove and
replace the protective shroud instead of the ball. The shroud
is removed from the end of the sensor block by unscrewing it.

5.1.3 Submersible Systems

The procedure is identical to that for the Dip System – see
Section 5.1.2 Dip Systems.

5.2 Replacing the Oxygen Sensor Block

5.2.1 Floating Ball System – Fig. 5.2
1) To avoid damaging the sensor capsule dry around it then
carefully remove it from the end of the sensor block and
store safely. DO NOT touch the membrane.
Caution. Do not leave the capsule exposed to air for more
than 30 minutes as the membrane will dry out.
2) Disconnect the cable from the dissolved oxygen transmitter
(or junction box if applicable) and attach a draw wire of
suitable length.
3) Slacken the cable gland on the end of the boom.
Securing Ring
Elbow Union
Boom
Ball Float
Sensor Block
Optional Water
Wash Nozzle
Sensor Capsule
Fig. 5.2 Replacing the Oxygen Sensor Body
4) Remove the self tap screw and pull off the end cap.
5) Draw the cable through the gland.
6) Unscrew the securing ring of the ball float (whilst restraining
the ball from turning) and withdraw the ball from the elbow
union.
7) Carefully unscrew the sensor block from the end of the
sensor pipe.
8) If fitted, disconnect the water wash feed tube from the
sensor block.
9) Pull the cable through the boom, leaving sufficient draw
wire at the end cap end to enable the replacement to be
pulled through.
10) Disconnect the sensor block cable from the draw wire.
11) Attach the draw wire to the replacement sensor block cable
and re-assemble in the reverse order of the previous steps.
12) Remove the dummy capsule from the replacement sensor
block and carefully replace the sensor capsule into the
sensor block.
5.2.2 Dip Systems
The procedure is identical to that for the floating ball system
except that it is only necessary to remove and replace the
protective shroud instead of the ball. Remove the shroud by
unscrewing it.
Optional Water Wash Feed Tube
Connecting Cable
Temperature Compensator
5 MAINTENANCE...
End Cap
Sensor Pipe