Camera settings
P
LEASE NOTE
The wizard leads you
step by step through the
calibration of the
shading correction.
The Dark correction
image is an acquisition
during which no light
falls on the camera. This
image is subtracted
from the acquired
image.
You will need to acquire new illumination correction images every time you make
changes in your optical system. For example, when you clean your microscope,
exchange the objectives or change the camera.
Acquiring images for the shading correction
1)
Open the
Camera Settings > Adjustment
ton.
The
Shading Correction
2)
If you have already acquired a Dark correction image, the
correction image
check box will be automatically selected.
Clear the check box, if you want to acquire a new Dark correction image.
Click the
Next >
button.
dialog box. Click the
wizard will open.
Calibrate...
but-
Skip recording dark