Olympus CK40M FRAME Руководство по ремонту - Страница 30
Просмотреть онлайн или скачать pdf Руководство по ремонту для Микроскоп Olympus CK40M FRAME. Olympus CK40M FRAME 38 страниц. Compact inverted metallurgical microscope

CK40M
5-2
Adjusting CK4M-MA illumination center
1)
Temporarily fix the stage. (Screw: AB5X15SA 4 pcs.)
2)
Loosen the screws (ANU3X4SA 2pcs.) fixing the slider mount.
3)
Insert slide glass between the slider mount and the revolving nosepiece and fix the
slider mount where it becomes flat.
4)
To look at the bulb filament, remove the frosted glass from the illuminator. (Remove
the screws: ANU 3X4SA 3pcs.)
5)
Using the centering telescope, focus on the filament.
6)
The positional relationship between the filament and the exit pupil should be as
shown below.
7)
Don't forget to mount the frosted glass on the illuminator.
The filament should be near
the center of the exit pupil.
C. REPAIR PROCEDURE
C-9
Exit pupil
Filament