Olympus SZX2-ILR66 Руководство по эксплуатации - Страница 12

Просмотреть онлайн или скачать pdf Руководство по эксплуатации для Лабораторное оборудование Olympus SZX2-ILR66. Olympus SZX2-ILR66 28 страниц. Led four-part ring illuminator system

5
³
|
@
²
Fig. 3
@
²
³
Fig. 4
9
Mounting on the STM6 Series (Fig. 3)
# When the MM6-VL/MM6C-VL vertical illumination arm is used with the
STM6 series, the SZX-LGR66 ring light guide adapter is required for
mounting.
1. Using the Allen wrench provided with the microscope, loosen the two
clamping screws ² on the SZX2-ILR66 illuminator @.
6
2. Insert the SZX-LGR66 ring light guide adapter ³ all the way into the SZX2-
ILR66 @ and tighten the clamping screws ² using the Allen wrench.
3. Loosen the clamping screw | on the SZX-LGR66 ³ using the Allen wrench.
4. Fir the SZX-LGR66 ³ into the objective 5 of the STM6, align the height of
the clamping screw | with the positioning groove 6, and tighten the
clamping screw | using the Allen wrench.
Mounting on the SZ61/51 Series (Fig. 4)
1. Fir the SZ-LGR66 ring guide adapter @ all the way into the auxiliary objec-
tive mount thread section at the front tip of the SZ61/51 microscopes.
}When an applicable auxiliary objective (110AL0.62X, 110AL0.75X or 110AL1.5X)
is mounted, attach the SZ-LGR66 to the front tip of the auxiliary objective.
2. Using the Allen wrench provided with the microscope, loosen the two
clamping screws ³ on the SZX2-ILR66 illuminator ².
3. Fit the SZX2-ILR66 ² all the way into the SZ-LGR66 @ and tighten the
clamping screws ³ using the Allen wrench.