Astrea Evolve Kullanıcı Kılavuzu - Sayfa 9
Endüstriyel Ekipmanlar Astrea Evolve için çevrimiçi göz atın veya pdf Kullanıcı Kılavuzu indirin. Astrea Evolve 19 sayfaları. 350mm column instructions for replacing the wetted flowpath using the column refresh kit
a)
Figure 8 Reassembly of Adjuster Assembly
9.
Move the adjuster assembly over the flowcell and lower so the flowcell assembly slides into the threaded height
adjust rod/seal setting ring assembly (see a) Figure 8) and ensure the adjuster seal is not trapped beneath the seal
setting ring (see b) Figure 8).
10.
Locate the spacer and fit the seal actuation nut (see c) Figure 8).
11.
Locate the flowtube seal and fit the sanitary clamp adapter.
12.
The refresh of the adjuster assembly is now complete. Ensure the unit is firmly seated on the floor and detach the
hoist.
®
Evolve
350mm Column: Instructions for Replacing the Wetted Flowpath using the Column Refresh Kit
b)
CAUTION!
Do not over tighten as this may lead to failure of the component
c)
9