Guidance and manufacturer's declaration - electromagnetic immunity
The KD-5904 is intended for use in the electromagnetic environment specified below. The
customer or the user of the KD-5904 should assure that it is used in such an environment.
IMMUNITY test
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration - electromagnetic immunity
The KD-5904 is intended for use in the electromagnetic environment specified below. The
customer or the user of the KD-5904 should assure that it is used in such an environment.
IMMUNITY test
IEC 60601test level
IEC 60601test level
Table 3
Compliance
17/19
Compliance level
± 6 kV contact
± 8 kV air
3 A/m
Electromagnetic environment
level
Portable and mobile RF
communications equipment
should be used no closer to any
part of the KD-5904, including
cables, than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance:
Operation Guide
Electromagnetic
environment -
guidance
Floors should be
wood, concrete
or ceramic tile. If
floors are
covered with
synthetic
material, the
relative humidity
should be at least
30 %.
Power frequency
magnetic fields
should be at
levels
characteristic of a
typical location in
a typical
commercial or
hospital
environment.
- guidance