Anritsu Company ShockLine MS46522B Series Hızlı Başlangıç Kılavuzu - Sayfa 10
Ölçüm Cihazları Anritsu Company ShockLine MS46522B Series için çevrimiçi göz atın veya pdf Hızlı Başlangıç Kılavuzu indirin. Anritsu Company ShockLine MS46522B Series 20 sayfaları. Vector network analyzers verification kits and performance verification software
6.
Adding Performance Verification Tools
This section describes how to enter the required testing tools and components into the PVS program to perform
a full performance verification.
Adding the PVS Tools
The following tools are required to perform a full VNA performance verification:
Verification Kit
• 50 ohm airline
• Beatty standard airline
• 20 dB Attenuator
• 50 dB Attenuator
Cal Kits
• Male
• Female
Refer to
Figure 7 on page
1. Select Utilities.
2. Select the drop down Asset Utility.
3. The Station Asset Utility menu will appear. See
Figure 9. Station Asset Utility Menu
4. Select the Import Verification Kit Data tab on lower left of the Station Asset Utility menu.
QSG-8. From the PVS Menu Bar:
Figure 9.