Great Plains Industries FM-300HR Kullanıcı El Kitabı - Sayfa 7

Ölçüm Cihazları Great Plains Industries FM-300HR için çevrimiçi göz atın veya pdf Kullanıcı El Kitabı indirin. Great Plains Industries FM-300HR 16 sayfaları. Electronic disc meter

Great Plains Industries FM-300HR Kullanıcı El Kitabı
Dispense-Display Field Calibration Procedure
Your Actions
1. Hold down CALIBRATE while pressing and
releasing DISPLAY until the Field Calibration
curve appears ("CAL C" message will be
displayed). Release both buttons.
2. To calibrate, press and hold the CALIBRATE
button. While continuing to hold CALIBRATE,
also press and hold the DISPLAY button. Hold
both buttons for about 3 seconds until you see
a blinking "dd CAL" message. Once the
"dd CAL" message appears, release both
buttons. You are now in field calibration mode.
3. Once the buttons have been released from
Step 2, the display will show the blinking
message "run 01."
4. If you want to continue with the calibration,
but have not dispensed any fluid yet, make
your final preparations to your pumping
system, but don't start pumping yet.
5. Start your pumping system so that fluid flows
through the meter. The display will stop
blinking. Dispense into a container that allows
you to judge the amount of fluid pumped.
When you have pumped the desired amount
(for example, 10 gallons), stop the fluid flow
quickly.
(continued on next page)
Notes
Remember that Field Calibra-
tion curves are not preset.
This step puts the unit in
dispense-display field cali-
bration mode ("dd CAL").
The computer is waiting for
you to make a decision to
either exit from field calibra-
tion mode or to begin a
dispense run. If you want to
exit the calibration now, go
to Step 11.
When the computer displays
a non-blinking "run 01" mes-
sage, it is sensing fluid flow.
For accurate results, dispense
at a flowrate which best simu-
lates your actual operating
conditions. Avoid "dribbling"
7