Optional modules
Commissioning the dispensing pumps
Before the liquid dispensing pumps can be
used the liquid agent must be drawn up into
the system and the amount to be dispensed
set.
Dispenser calibration
The dispensers need to be calibrated to
ensure the dispenser pumps supply the
correct amount of liquid agent.
To calibrate the dispensers with the machine
switched off:
^ Press and hold in button 2.
^ Then switch the machine on with the "On"
switch.
Dispenser calibration ~ Main menu
Dispenser pump 1
Dispenser pump 2
Selection choice: Dispenser pumps 1 to 13
After selecting a dispenser pump the
following message will appear in the display:
Dispenser calibration ~ Dispenser pump 1
V: 100ml
t: 0s
The amount to dispense can be set between
0 and 9999 ml in 1 ml increments.
^ Press the 0/1 multi-function button and the
dispenser hose will start to fill. Press the
0/1 button again to stop filling the hose.
^ Direct the hose of the dispenser pump
being calibrated into a suitable measuring
flask.
,
Avoid contact with the skin and eyes,
and wear suitable protective clothing.
^ Press the A "Back" button to return to the
main menu.
28
e
A
no
A
S
0/1
Dispenser calibration ~ Main menu
Dispenser pump 1
Dispenser pump 2
^ Select dispenser pump.
Dispenser calibration ~ Dispenser pump 1
V: 100ml
t: 0s
^ Press button 0/1.
To stop the pump press the 0/1 button
again, or it will switch off automatically
after a period of 60 seconds.
^ Measure the amount collected in the
measuring flask.
Dispenser calibration ~ Dispenser pump 1
V: 100ml
t: 30s
^ Enter this value in field "V:". Flow rate P in
ml/min will be calculated electronically.
^ Press the S "Save" button to save the
values measured in memory.
^ Reconnect the hose.
Cleaning and maintaining the dispensing
system
The dispensing system should be flushed
through with clean warm water every two
weeks and before any longer down-times to
prevent the risk of blockages and corrosion.
^ The suction pipes should be cleaned
through with water.
^ Place the suction pipes in a container of
hot water (40 - 50°C).
^ Then use the dispenser calibration system
to operate each pump until the dispensing
system is thoroughly flushed through.
^ After reconnecting all parts check all
connections, hoses, dispensers and seals
for leakages.
e
A
no
A
S
0/1
P: 200.0 ml/min
A
0/1
S