AQUARION AQ111 Bakım Kılavuzu - Sayfa 2
Tıbbi Ekipman AQUARION AQ111 için çevrimiçi göz atın veya pdf Bakım Kılavuzu indirin. AQUARION AQ111 4 sayfaları. Self-contained eyewash station
4.
Removed the tamper evidence seal, compress the white push button and then remove the plug/cap.
Figure 4: Seal removed, button compressed and cap removed.
5.
Place the funnel in the open valve and press down gently to secure it in place.
6.
Fill the bag with 4.0-4.2 gallons of deionized potable water (15.1-15.9 L).
Note: Filling the cartridge with the correct amount of water is critical to the system meeting the ANSI Z358.1
Standard.
Doc. No.: 01999986
Revision: D
Issue Date: 08/15/21
®
Figure 3: Aquarion
valve slid into slot.
2 of 4