Olympus UHI-2 Bakım Kılavuzu - Sayfa 15

Tıbbi Ekipman Olympus UHI-2 için çevrimiçi göz atın veya pdf Bakım Kılavuzu indirin. Olympus UHI-2 38 sayfaları.

Olympus UHI-2 Bakım Kılavuzu
1-2 The indicator is not lit. The panel switch is not accepted. (Error check
Error check
Check the power supply procedure.
Does the panel
work?
End
Check the connection.
Operate?
Check the power supply U.
Operate?
Replace the UPUH25U.
Operate?
Replace the front panel or panel sheet.
Operate?
Replace the UPUH15U.
2-3
TROUBLESHOOTING
Check the connection of all harness.
Especially, the connection of the
flat cable between UPUH25U and
panel sheet.
Check the voltage at J8 of UPUH15U.
DP0591
DP0797
(If the voltage is out of value, replace DP0591 or
DP0797.)