Display
Meaning/Action
If the unit displays 'Err2' after Step
5, it means the excessive saliva went
into the sensor during the test. Wait
20 minutes and then try again.
When the total number of tests is over
500 times, the recalibration service
is needed. Please send the device
to ACE.
If the unit displays 'CAL' after
countdown, it means the tester has
been used over 365 days from its last
calibration. Please send the device for
a calibration service to ACE.
Remaining days to recalibration,
e.g. d365 means 365 days left to
recalibration. Contact ACE for proper
recalibration service.
The elapsed day from the required
recalibration date, which is a reminder
to do the recalibration, e.g. d.001
means 1 day elapsed from the requi-
red recalibration date. Please send
the device for a calibration service
to ACE.

Technical data

Statement of the BAC
0.07 ~ 5.00 ‰ (fixed basic setting),
under 0.07 ‰ the test result is shown
as 0.00 ‰
Accuracy
+/- 0.05 ‰ at 0.5 ‰
Warm-up time
15 seconds up to max. 2 minutes
Response time
Within 12 seconds
Sensor
Fuel cell sensor
Energy supply
2 pieces AAA alkaline batteries (1.5 V)
Battery capacity
Approx. 1.000 tests
Dimensions of the device
47 mm x 108 mm x 17 mm (W x H x D)
Weight
85 g (including batteries)
Calibration
Every 6-12 months or after approx.
500 tests
Ambient conditions
Operation: 5 - 40° C
Storage: 0 - 50° C