APPLY STABLELOC FIXATOR
Making sure the distal clamp is loose (1) and the
main body set screws are loose (2), slide the
Stableloc Fixator Assembly onto the Fixator Pins. Align the
Stableloc Fixator on the wrist so the ball joint (3) is positioned
over the center of wrist rotation. This will aid in the accuracy of
subsequent adjustments.
Lightly tighten the distal pin clamp, the ball shaft locking
screw, the ball joint set screw and the main body set screws,
provisionally locking the Stableloc Fixator into position.
5
DISTRACTION/LIGAMENTOTAXIS
To apply distraction to the wrist, turn the distraction
nut towards the housing to apply the required
amount of distraction. One complete revolution of the
distraction nut equals 1 mm of distraction (1). Lock the ball
shaft locking screw (2).
Note:
The distraction nut is marked with "A,B,C,D" as a
reference to indicate where distraction began.
6
DORSAL/PALMAR TRANSLATION
Place the Stableloc Hex Driver into the jack screw
(1) on the dorsal side of the main body of the
fixator. Turn the Stableloc Hex Driver clockwise for palmar
translation and counterclockwise for dorsal translation.
The surgeon is able to determine the amount of translation by
the scale on the side of the Stableloc Fixator. There are 8 mm
of translation available in either direction.
7
RADIAL/ULNAR DEVIATION
Gain proper anatomic alignment of the fracture on
the radial/ulnar plane by loosening the distal pin
clamp (1). This allows free movement in both the medial/lateral
(M/L) and radial/ulnar direction.
Note:
Traction may be lost when the distal pin clamp locking
screw is loosened.