Abbott i-STAT 1 Kullanıcı Kılavuzu - Sayfa 2
Ölçüm Cihazları Abbott i-STAT 1 için çevrimiçi göz atın veya pdf Kullanıcı Kılavuzu indirin. Abbott i-STAT 1 9 sayfaları. For use with a clia certificate of waiver
Ayrıca Abbott i-STAT 1 için: Hızlı Referans Kılavuzu (7 sayfalar)
System Components
i-STAT 1 Analyzer
Downloader
Downloader/Recharger
For in vitro diagnostic use.
See System Manual for instructions.
Intended Use: The i-STAT 1 Analyzer is intended for use with i-STAT cartridges
for the in vitro quantification of various analytes in whole blood.
2
i-STAT Cartridges
Electronic
Simulator
i-STAT 1
i-STAT 1
i-STAT Printer
Martel Printer
Quality Check Messages and Codes
Message
Cause
Date Invalid,
Date outside six
Check Clock
month lifetime of
software.
Dead Batteries,
Insufficient power to
Replace Batteries
complete a test cycle.
Temperature Out of
Temperature outside
Range, Check
operating range of 16
Status page
to 30 °C.
Invalid or Expired
Software expired or
CLEW
corrupt.
Analyzer Interrupted,
Last cartridge run not
Use Another
completed.
Cartridge
Cartridge Error
Usually problem with
sample or cartridge
filling.
Cartridge Preburst
Calibrant pack burst
before cartridge inserted
into handheld.
Unable to Position
Cartridge not sealed.
Sample
Clot in sample.
Aberrant cartridge.
Sample Positioned
Cartridge underfilled.
Short of Fill Mark
Sample Positioned
Cartridge overfilled.
Beyond Fill Mark
Test Cancelled by
User did not respond
Operator
to mandatory prompt
before handheld time
out.
Cartridge Type Not
Software does not
Recognized
recognize cartridge.
Action
Select 5-Clock Set from
Administration Menu. (Password
protected.)
Replace disposable batteries or
recharge the rechargeable battery.
Check handheld temperature by
pressing 1 for Analyzer Status under
the Administration Menu. Move
handheld to warmer area if below
operating range or to cooler area if
above the range.
Verify that the handheld's date is
correct. Change software if expired.
Update again if not expired.
Check the battery pack inserted
properly. Check for Low Battery
startup warning.
Use another cartridge. If same code
repeats more than twice, try another
handheld.
Use another cartridge - do not press
on center of cartridge. Check that
cartridges have not been frozen.
Use another cartridge.
Use another cartridge - fill to Fill
Mark.
Use another cartridge - do not fill
beyond Fill Mark.
No action required.
Update software. Check to see if
cartridges are expired.
15