Apogee Instruments S2-432 Посібник користувача - Сторінка 10

Переглянути онлайн або завантажити pdf Посібник користувача для Аксесуари Apogee Instruments S2-432. Apogee Instruments S2-432 18 сторінок. Red-far-red sensors

Apogee Instruments S2-432 Посібник користувача

OPERATION AND MEASUREMENT

The S2-432 Red - Far-red sensor has a Modbus output, where red and far-red photon flux density are returned in
digital format. Measurement of S2-432 Red - Far-red sensors requires a measurement device with a Modbus
interface that supports the Read Holding Registers (0x03) function.
Wiring
The Green wire should be connected to Ground to enable RS-485 communication, or it should be connected to 12
V power for RS-232 communication. Text for the White and Blue wires above refers to the port that the wires
should be connected to.
Sensor Calibration
All Apogee Modbus Red - Far-red sensors (model S2-432) have sensor-specific calibration coefficients determined
during the custom calibration process. Coefficients are programmed into the sensors at the factory.
Modbus Interface
The following is a brief explanation of the Modbus protocol instructions used in Apogee S2-432 Red - Far-red
sensors. For questions on the implementation of this protocol, please refer to the official serial line
implementation of the Modbus protocol:
(2006) and the general Modbus protocol specification:
http://www.modbus.org/docs/Modbus_Application_Protocol_V1_1b3.pdf
found at:
http://www.modbus.org/specs.php
Overview
The primary idea of the Modbus interface is that each sensor exists at an address and appears as a table of values.
These values are called Registers. Each value in the table has an associated index, and that index is used to identify
which value in the table is being accessed.
Sensor addresses
Each sensor is given an address from 1 to 247. Apogee sensors are shipped with a default address of 1. If using
multiple sensors on the same Modbus line, the sensor's address will have to be changed by writing the Slave
Address register.
http://www.modbus.org/docs/Modbus_over_serial_line_V1_02.pdf
White: RS-232 RX / RS-485 Positive
Blue: RS-232 TX / RS-485 Negative
Green: Select (Switch between RS-232 and RS-485)
Black: Ground
Red: Power +12 V
(2012). Further information can be