Apogee Instruments SI-400 Series Посібник користувача - Сторінка 12

Переглянути онлайн або завантажити pdf Посібник користувача для Вимірювальні прилади Apogee Instruments SI-400 Series. Apogee Instruments SI-400 Series 20 сторінок. Infrared radiometer

Apogee Instruments SI-400 Series Посібник користувача
Sensor Calibration
Apogee SI series infrared radiometers are calibrated in a temperature controlled chamber that houses a custom-
built blackbody cone (target) for the radiation source. During calibration, infrared radiometers (detectors) are held
in a fixture at the opening of the blackbody cone, but are thermally insulated from the cone. Detector and target
temperature are controlled independently. At each calibration set point, detectors are held at a constant
temperature while the blackbody cone is controlled at temperatures below (12 C), above (18 C), and equal to the
detector temperature. The range of detector temperatures is -15 C to 45 C (set points at 10 C increments). Data
are collected at each detector temperature set point, after detectors and target reach constant temperatures.
All Apogee analog infrared radiometer models (SI-100 series) have sensor-specific calibration coefficients
determined during the custom calibration process. Unique coefficients for each sensor are provided on a
coefficient certificate (example shown below).
Calibration overview data are listed in box in upper left-hand corner, sensor specific calibration coefficients
are listed in box in upper right-hand corner, temperature errors are shown in graph, and calibration date is
listed below descriptions of calibration procedure and traceability.