ADE M600020 Посібник з експлуатації - Сторінка 9
Переглянути онлайн або завантажити pdf Посібник з експлуатації для Ваги ADE M600020. ADE M600020 16 сторінок. Electronic bed scale
Determining the Body Mass Index (BMI)
The BMI figure is an index showing the ratio between height and weight.
body weight
BMI=
(height)²
display. Use the numerical key pad to enter a new value for the height (in cm) or retain the value, which is
already been displayed. Confirm the value entered with the ENTER key or delete same by using the CLEAR
key. After the entry has been confirmed by the ENTER key, the BMI value will be calculated using the weight
on the scales. The display changes between weight value and BMI value "b xx.x" (every two seconds) for a
period of 3 minutes. Press the BMI key (11) to cancel before this time has expired.
Evaluating Body Mass Index
BMI evaluation according to WHO
BMI
<
16
16
-
17
17
-
18.5
BMI
18.5
-
25
BMI
25
-
30
30
-
35
35
-
40
>
40
Print out a record
The scales must be at standstill, e.g. standstill status display
Press the ENTER key (13) to send the data to the serial interface.
Note: Printing is not possible if the net weight is 0. The print function only produces one print-out. A reset is
required to print out another copy.
Cleaning
Remove the mains plug before cleaning the unit. Do not use any aggressive cleaning materials for cleaning;
only use a damp cloth or conventional disinfectant. Ensure that no liquids or any fluid gets into the scale.
Error messages
Error mes-
sage
Overload display
Overflow indicator (negative)
Overflow indicator (positive)
AD LOW
A/D above negative range
AD HI
A/D above positive range
M600020-210322-Rev008-UM-e
To determine this value you need the height of the person.
Start the scale and weigh the person following the steps under weighing. Then
press the BMI key (11) to start BMI mode. The last value entered appears on the
Severely underweight
Moderately underweight
Slightly underweight
Normal weight
Overweight
Obese Class I
Obese Class II
Obese Class III
Care, servicing and recalibration
Description
(14) appears at the screen.
Problem solution
Weight exceeds max. capacity of the
scales.
Negative weight indicated – weight too
low to be displayed (<=99999).
Positive weight indicated – weight too
high to be displayed (>=999999).
Check if the scales are correctly as-
sembled or if they are damaged.
Check if the scales are correctly as-
sembled or if they are damaged.
9