AUTOSCALER AW30C Посібник - Сторінка 11
Переглянути онлайн або завантажити pdf Посібник для Конвертер мультимедіа AUTOSCALER AW30C. AUTOSCALER AW30C 14 сторінок.
source without a purge cycle. The scaler air line will bleed off to
zero pressure protecting the AUTOSCALER®.
3. Plug the power cord into the wall (see electrical requirements).
4. Select an insert. Moisten the rubber O-RING before inserting into
the handpiece. While holding the handpiece, UPRIGHT, place the
insert in the open end of the handpiece, and with a gentle PRESS
and TURN motion, snap the insert in fully. DO NOT force the insert.
5. Manufacturer recommends the use of a disposable handpiece
sleeve/barrier be used over the handpiece assembly. Place the
sleeve/barrier over the handpiece. While holding the sleeve/
barrier in place, puncture the insert through the plastic at the
open end of the handpiece, and snap the insert in fully. Dispose of
sleeve/barrier after each patient.
6. Due to atmospheric exposure, and exposure to sterilization
solutions, the O-RING on the insert tip may become swollen,
brittle, or crack, and require replacement.
7. Set the POWER KNOB all the way to the LEFT in the LOW position
8. Hold the handpiece in a VERTICAL upright position, and step on
the foot switch. Within 10-15 seconds water should come out of
the open end of the handpiece IF NOT, QUICKLY
turn WATER KNOB clockwise 1-2 rotations to get
water flow. Some air will escape at the same
time. This step is IMPORTANT because trapped
air will cause HOT SPOTS (usually noted by a
loud screeching sound) which can affect the
performance of the AUTOSCALER® or
permanently damage the handpiece assembly.
9. Inserts and handpiece will burn in 45-60 seconds in the absence
of water. A (18cc—30cc/+/min) flow is the normal flow range, and
necessary at all times. NEVER touch the Cavitating/Vibrating end
of the insert tip to any flesh such as the cheek, tongue, or gums
unless performing special periodontal procedures. This will cause a
FRICTION BURN to the patient. Always keep the insert in motion.
Do not rest the handle or insert on the lips during a procedure.
The insert tip should touch primarily the calculus on the tooth
surface, and NOT the enamel of the tooth surface.
10. Allow the air bubbles to escape for about 10-20 seconds. Turn the
POWER KNOB to the RIGHT in the Medium position, and the
insert tip should Cavitate/Vibrate with a strong mist. Turn the
water knob to the right if there is no mist or weak misting.
21
11. The application of the power knob is for controlling the Output
Power to the insert tip in a smooth linear action. The operator can
select the EXACT power needed for cleaning. Set power knob LOW
(Left Direction) to MEDIUM (Center) for cooler handpiece/insert
temperature.
12. The high power setting is for applications where an excessive
amount of calculus is built up. The Low power setting can be used
for periodontal Subgingival curettage and root planing
procedures. Most scaling should be done only on the LOW to
MEDIUM power setting.
13. If SLIMLINE™, FSI™ or Modified Periodontal inserts are used, set
the POWER Knob in the BLUE shaded area. This power level is
designed to be used with these fragile specialty inserts. Inserts
may give a drip and mist combination, this is normal for these
inserts. This is actually preferred by most Periodontal
professionals due to the water drip being ultrasonically activated
within 0.5mm of the tip assisting in killing plaque.
UNIT IS NOW READY FOR USE
SHUT DOWN PROCEDURE
Be sure to shut off Air and Water supply to the scaler when
unattended. The installing service technician will indicate to the
operator which controls perform this action.
TROUBLESHOOTING
If the unit does not function, proceed with the following:
1. Check the plug. Is it all the way in the receptacle?
2. Try another insert. Make sure it is the correct frequency 25K or
30K depending on your model. Each model has a frequency (25K or
30K) label on the cover and the frequency is also indicated on the
serial number plate located on the bottom of the device. The most
common mistake when purchasing a new unit is selecting the
wrong frequency insert.
3. If water flow is insufficient, check water pressure at the operatory for
minimum of 20 P.S.I. The Insert Water slot could be clogged with
debris. Try another insert.
4. If the handpiece handle or insert operates HOT: REDUCE POWER
LEVEL and INCREASE WATER FLOW. Depending on the water
regulator supplied in your operatory the following applies: HIGHER
water flow to the RIGHT, and LOWER water flow to the LEFT. Some
water regulators adjust the opposite (more flow to left and less to the
22