Olympus BXC-FSU Інструкція з експлуатації - Сторінка 22
Переглянути онлайн або завантажити pdf Інструкція з експлуатації для Мікроскоп Olympus BXC-FSU. Olympus BXC-FSU 41 сторінка. Modular microscope assemblies hardware
2.
Loosen the mounting screw (C) of the light guide with the Allen screwdriver provided with BXC-RLI-LGCA and attach
the light source using light guide.
3.10.6 Attaching a third-party light source
You can attach a third party light source instead of the light source explained in 3.10.1 to 3.10.3 and at *1 shown in
3.10.4 or *2 shown in 3.10.5 .
NOTE:
Product performance and safety cannot be guaranteed if a third-party light source is used with the product. If you wish to
use this product in combination with a third-party light source, take measures under your responsibility so that the use of
this product in combination with systems you will use complies with standards required in your local area.
3.11 Setting up BXC-FSU
3.11.1 Product outline of BXC-FSU
BXC-FSU is an intermediate mounting unit that is mounted on the top of the illuminator (BXC-RLI/BXC-RLI-CA/BXC-RLI-
LGCA). It detects the focus state of the sample and connects to the control box (BXC-CBB, BXC-CBE1) to output Focus Error
Signals and Focus Status Judgement Signals (In-Capture signal, In-Focus signal).
The following figure is an example of a typical system with BXC-FSU installed. The customer can control the Z-axis drive
mechanism prepared by the customer using the Focus Error Signal and Focus State Judgement Signal (In-Capture signal,
In-Focus signal). Also, when the safety lock signal is OPEN, the LD light emission in the BXC-FSU can be blocked.
3. Setup procedures
18
3.11 Setting up BXC-FSU