(4) Safety Lock
Label: S-LOCK
Function: Laser is not emitted if not terminated.
Relating the safety lock terminal status to outer panels of the customer's device, Laser radiation can be avoided when the
panels are open.
Connector type: BNC
Terminal Condition
OPEN
Connect to GND (SHORT) Un-Lock (Laser enable)
* Return from "Locked condition": Connect "Terminal" to GND again. No need to reboot BXC-CBB.
3.11.3 Adjusting camera diopter
Adjust camera diopter using the following steps.
TIP:
Be sure to adjust camera diopter, to prevent the following issues.
1. Parfocal among objective lenses getting worse.
2. Auto-focusing performance getting worse.
1.
Put a specimen on the stage.
2.
Set the highest magnification objective lens in the light path.
Command: 1OB
3.
Move the Z-stage and adjust focus on the specimen.
4.
Set the lowest magnification objective lens in the light path. Be sure not to move the stage height.
Command: 1OB
5.
Move the circular dovetail of the camera adapter and adjust focus on the specimen.
21
3. Setup procedures
Laser Condition
Lock (Laser disable)
3.11 Setting up BXC-FSU