Olympus CH20 Посібник з ремонту - Сторінка 10
Переглянути онлайн або завантажити pdf Посібник з ремонту для Мікроскоп Olympus CH20. Olympus CH20 31 сторінка. Biological microscope
CH10/CH20
1-2 Observation tube
Part
Item
Binocular
Interpupillary
tube
distance
adjustment
range
Interpupillary
distance
working force
Diopter
Compensa-
difference
tion range
compensa-
tion
Optical
Left/right
axis
optical axis
Absolute
optical axis
(right sleeve
only)
Exit pupil
center
Tube length Absolute
parfocality
B. INSPECTION STANDARD
Standard
53 mm or less at
min. - 75 mm or
more at max.
3.9 - 9.8N {400 -
1000g}
±5 diopter
On image surface:
0.15 mm or less in
vertical direction
0.15 mm or less in
outward direction
0.3 mm or less in
inward direction
0.3 mm or less on
image surface
Within 20% of
objective's exit
pupil diameter
±0.5 mm or less on
image surface
B-2
Method
Set the observation state, insert a thin paper with
graduations at the eye point position and measure
the interpupillary distance.
Tie a string around the sleeve periphery, work the
interpupillary distance to measure the working
force with a tension gauge.
Turn the diopter ring and check that the scale
goes over the limit.
Observe a specimen whose center can be
identified (e.g., on concentric circles) using the
standard eyepiece (KN0048; with adapter-1) and
4X or 10X objective.
Align the center of specimen with that of visual
field in the right sleeve as the standard, and read
the displacement between centers of the
specimen and the visual field in the left sleeve
using the reticle scale (1 graduation is 0.1 mm.)
of KN0048.
Set the interpupillary distance to 62 mm, combine
the standard eyepiece (KN0048; with adapter-1),
the microscope frame (product) and the standard
objective (KN0015), and read the displacement
between centers of the specimen and the visual
field in the right sleeve using the reticle scale of
KN0048.
Combine the centering telescope (KN0029), 4X
or 10X objective, and the microscope frame
(product), and read the displacement between the
objective's exit pupil diameter center and the
cross hairs center of centering telescope.
Combine the standard eyepiece (KN0048; with
adapter-1), the focusing telescope (FT-36), the
microscope frame (product), and the standard
objective (KN0016).
Set the interpupillary distance to 62 mm, and read
the parfocality difference between the reticle in
KN0048 (with adapter-1) and the specimen in
KN0016 using helicoid scale (1 graduation is 0.1
mm.) of KN0048.