Acqiris U1084A Посібник із запуску - Сторінка 18

Переглянути онлайн або завантажити pdf Посібник із запуску для PCI карта Acqiris U1084A. Acqiris U1084A 20 сторінок. Acquisition card, 8-bit, up to 4 gs/s and 1.5 ghz bandwidth

Step 5: Using the Applications

Switching firmware

The U1084A digitizer can be purchased with different firmware options.
This section contains important information if your module as been order with more than one of
the following firmware:

Context

When your module has two or more firmware options, only the one stored in the default
EEPROM memory can be used.
If you want to use another firmware option, you have to switch firmware.

Default Firmware

The Table below gives the firmware loaded by default.
U1084A00x Options
DGT,DGS,AVG,TDC
DGT,DGS,AVG
DGT,DGS,TDC
DGT,DGS
DGT,AVG
DGT,TDC
AVG,TDC
DGT

Switching firmware process

Follow the instructions below in order to switch the firmware:
1. From C:\Program Files (x86)\Agilent\MD1\, launch the application AcqEEPROMProg.exe.
2. Then, depending on the option you want to use:
3. Then, select the Averager, Digitizer or PeakTDC firmware function in the Firmware
4. The update of the module starts and could take some time (~15 minutes). DO NOT
18
Digitizer firmware: used for DGT (Digitizer), or DGS (Digitizer with simultaneous acquis-
ition and readout (SAR)) options.
Averager firmware: for AVG (Real-time Averaging) option,
PeakTDC firmware: for TDC (Time-to-Digital Conversion and Peak Analysis) option.
a. To load the averager firmware : select C:\Program Files (x86)\IVI Found-
ation\IVI\Drivers\AgMD1\Firmware\u1084_ave.vme
b. To load the digitizer firmware: select C:\Program Files (x86)\IVI Found-
ation\IVI\Drivers\AgMD1\Firmware\u1084_dig.vme
c. To load the PeakTDC firmware: select C:\Program Files (x86)\IVI Found-
ation\IVI\Drivers\AgMD1\Firmware\u1084_tdc.vme
function dialog box.
STOP THE PROGRAM, it could damage your module. The end of the module update is
Default firmware
Averager
Averager
PeakTDC
Digitizer
Averager
PeakTDC
Averager
Digitizer
Acqiris U1084A Startup Guide