AMD AMD-3700 Installations- und Betriebshandbuch - Seite 6

Blättern Sie online oder laden Sie pdf Installations- und Betriebshandbuch für Medizinische Ausrüstung AMD AMD-3700 herunter. AMD AMD-3700 7 Seiten. Telephonic stethoscope

AMD AMD-3700 Installations- und Betriebshandbuch
B.

RX Mode

Without a Chest Piece Assembly plugged into the Chest Piece jack of the TR-1, the unit
is automatically in the Receive Mode.
established and a Headset plugged into the Headset jack of the unit, the clinician is set
up to hear the stethoscope sounds from the remote AMD-3700 (in Transmit Mode).
Volume to the Headset can be increased by turning the Volume control clockwise.
Increasing the volume too high can cause distortion in the Headset and will also
increase the background noise. Start with the lowest volume necessary to hear the
sounds, then turn up the volume as needed to focus in on specific sounds.
The B/D Switch can be used in an auscultation exam similarly to a traditional
stethoscope. The Bell (B) position cuts off the high end at 250 Hz and boosts the low
frequency sounds below 100 Hz for enhanced heart auscultation. The Diaphragm (D)
position slightly reduces the very low frequencies and extends the high frequencies to
1,400 Hz.
IV.

Maintenance and Calibration

There is no scheduled maintenance required for the AMD-3700 Telephonic
Stethoscope.
There is no calibration required for the AMD-3700 Telephonic Stethoscope.
If problems should arise with the AMD-3700 Telephonic Stethoscope, the table in the
following section may be used to trouble shoot the problem. Any failed units should be
returned for repair.
The AMD-3700 Telephonic Stethoscope may be cleaned with a damp cloth.
AMD Global Telemedicine, Inc.
73 Princeton Street North Chelmsford, MA 01863
978-937-9021 (phone) 978-937-5249 (fax)
With the Data Communications Channel
6