Calibration

1. Flow certified 21% Oxygen calibration gas into the sensor port on the front of
the monitor. Wait 3-5 minutes before starting calibration, then execute the
"SPAn" calibration.
(Continue to flow calibration gas through entire process)
2. Press Mode until CALI is flashing in small print under the O2 reading.
3. Press Enter to view the calibration settings.
4. "SPAn" and "O2" will be flashing.
5. Press Enter again to begin the calibration. "CALIBRATING" will begin flashing.
(Continue to flow calibration gas throughout eFntire process)
6. After approximately 3 minutes, the display either "PASS" or "FAIL".
7. If "PASS" press enter to save.
8. If "FAIL" repeat the process.
9. If "FAIL" more than twice, contact CO2Meter technical support.