Aktek ALIT-TIF Benutzerhandbuch - Seite 12
Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Messgeräte Aktek ALIT-TIF herunter. Aktek ALIT-TIF 16 Seiten. Ultrasonic level meter
Aktek Endüstriyel Ekipman ve Enstrümantasyon Ltd. Şti
3.3 The load resistance
Load resistance and supply voltage diagram
3.4 The cable
A cable comprising screened, twisted pairs is recommended, the diameter
should be between 5-9mm, the distributed capacitance of the cable should
be less than 0.0395uF, distributed inductance should be less than 0.5mH.
The cable shield must be grounded.
3.5 The explosion-proof connection
The explosion-proof type is intrinsically safe, the grade is ExiaIICT6. The instrument should be
connected with barrier when using. S2022 I.S. barrier is the associated apparatus of the
instrument, and its explosion-proof grade is ExiaIIC, power supply is DC24V (±10%).
www..aktek.com.tr
11