13.1 Refill FlexRx
To REFILL FlexRx on the DT controller, follow the instructions below:
1. Enter your Key Code to unlock the controller.
2. Device Code 'Cxyz' is displayed on screen
Note: If Device Code is not displayed, there are exposures remaining and it will be
displayed once all exposures have been used. Device Code can be viewed in Menu 5
prior to all exposures being used.
3. Contact prescribing physician and provide Device Code 'Cxyz'
4. Physician will contact Daavlin to authorize FlexRX refill
5. Daavlin Service will provide Treatment Refill Code to doctor or patient
14.0 Special Notes
1. The maximum time that can be entered is 59 minutes and 59 seconds.
2. It is not possible to enter XX:60 seconds. You must enter 1:00 minute instead.
3. If power goes off during a treatment, the system will not remember how much of the
treatment has elapsed. When power is restored, the controller will prompt for the Key Code
and another treatment will have to be started.
a. If Flex Rx is activated and the controller loses power during a treatment, an
exposure/full treatment will be assumed, and the remaining treatments will
decrease by one (1) treatment.
4. If Flex Rx is active, the display will flash the number of treatments remaining two (2) times
after a treatment or at power up if the amount of remaining exposures is twenty (20) or
less, signaling the user to obtain a new Flex Rx Treatment code soon.
5. Per the Daavlin Flex Rx Algorithm, there are 1000 Device Codes (C000-C999), each
having five (5) four-digit Flex Rx Treatment Codes (0000-9999) that set a predetermined
amount of exposures to be allowed by the controller.
a. 0 exposures (locked), 75 exposures, 100 exposures, 250 exposures, or unlimited
exposures (unlocked).
15.0 Care of the Unit
15.1 Recommended Maintenance Schedule
Item / Action
Clean all patient contact surfaces
Dusting of the unit and lamps
Fully clean all internal reflectors,
lamps, and protective acrylic
MNL-00049 [1]
Frequency
Between each treatment
Once a month
Annually (behind the lamps)
Proprietary and Confidential. Printed copies will not be updated.