Daavlin 7 Series Betriebshandbuch - Seite 5

Blättern Sie online oder laden Sie pdf Betriebshandbuch für Lantern Daavlin 7 Series herunter. Daavlin 7 Series 13 Seiten. Integrated dosimetry with link control
Auch für Daavlin 7 Series: Installationshandbuch (2 seiten)

Daavlin 7 Series Betriebshandbuch
7. Gently slide the unit into its permanent location.
8. Install the power cord by inserting the male "D" shaped plug into the "D" shaped receptacle at the back of
the device. Then, plug the unit into any grounded three prong electrical receptacle.
9. Insert the perpendicular grid wires into the two holes in the top lamp plate and push the grid all the way up.
Swing the bottom of the grid in toward the lamps and insert the grid wires into the two holes in the bottom
lamp plate. Then, lower the grid until the bracket touches the bottom lamp plate. Using a ¼" hex
driver/socket, screw the bracket to the bottom lamp plate using one of the hex head, self-tapping screws
that are provided.
Inspect the Lamps in Your Unit
The 7 Series can be prescribed with a variety of different lamps, each having a different effect on the skin. It is
important to check that the proper kind of lamps are installed in the device. Consult the Lamp Specification
Guide and if there is any question as to whether or not it is equipped with the lamps that you have been
prescribed, contact Daavlin immediately. The code numbers shown are generally located at the base of the
lamp.
Lamp Specification Guide
-F72T12BL/HO
-060OF72HO BL

Precautions

To avoid eye injury during treatment, you must wear the tightly fitting ultraviolet blocking goggles provided
with your device. Always use Daavlin approved eye wear.
If psoralens (photosensitizing drugs) are being used as part of your treatment ("PUVA"), your eyes should
be protected from exposure to ultraviolet (sunlight) for 24 hours after taking the drug.
A burn can result in as little as 15 seconds of exposure to UVB light. Prior to using your home phototherapy
unit contact your prescribing physician for specific treatment instructions and dosing information.
All other persons should vacate the treatment area during a treatment session to avoid unnecessary
exposure to light energy.
Consult your physician about protecting other areas of your body that may never have been exposed to
sunlight.
MANU0140
Rev. 0
2/2019
UVA
or
-FS72T12/NBUVB/HO/RDC/100W
Narrow Band
UVB
-TL-01/100W-FS72
or
5
Broad Band
UVB
-FS72T12HO