(Currently the value is indicated.)
2) Zero and Span Setting
Press Up key about 3 seconds to return to Real-weight Calibration mode.
->
Leave the sensor free-load and press ENT key to set Zero.
->
Press Up key to be
Press ENT key, It uses
value. (Standard weight value it was prepared).
Press ENT key
Press ENT key to set SPAN value.
->
If you want to get out of the setting mode and return to the measurement mode,
you may have to push the Function key.
※ If, when
is indicated or with Real-weight value different case, to repeat the
above statement.
3) Shunt CAL data
After finishing Real-weight Calibration, eliminate standard weight
and make display " 0 " .
When indicator displays "0", you turn on S.CAL S/W in back panel.
this then, indicator displays Shunt CAL data in front panel.
Please record Shunt CAL data .
Date
※ note)Please turn off SHOUT CAL S/W after recording Shunt CAL data.
4) Shunt CAL data calibration
Press Up key about 3 seconds to return to Real-weight Calibration mode.
->
Leave the sensor free-load and press ENT key to set Zero.
->
Press Up key to be
Press ENT key, It uses
->
->
Shift key and
, put counterweight on.
->
->
S.CAL data
->
->
Shift key and
12
Up Key and it inputs Real-weight
Up Key and it inputs Shunt CAL data.