5 Trouble shooting
If the suggested actions do not correct the
fault, call Amersham Pharmacia Biotech.
Fault
Action
Leakage
1 Check the tubing and connectors.
If leakage is located to the mixer
chamber, replace the complete
chamber.
Waves on the gradient
1 Check that the pump is operating
and is programmed correctly.
2 Clean the mixing chamber to make
sure it is free of dirt or particles.
3 Change to a larger mixing chamber
volume if necessary.
4 Run the mixer test from
UNICORN™ to determine the
working status of the mixer. If the
test failed, contact Amersham
Pharmacia Biotech.
Unlinear gradients or slow response to
%B changes
1 Check that the tubing has been
washed properly and that the pump
is operating.
2 Change to a smaller mixer volume.
Not running
1 Check cables.
2 Check Setup menu on the pump.
Other faults
Contact Amersham Pharmacia
Biotech.
4
Reference information
Description
A mixer motor
inside the housing
spins a magnet at
600 rpm, that
causes the stirrer
in the mixing
chamber to rotate.
The eluents are mixed in two steps:
1 Premixing in a static mixer with a small
volume (22 µl).
2 Dynamic mixing chamber with rotating
stirrer.
The mixer housing and the mixer chamber
contains no user replaceable items.
2
1
0 0 , , 6 6
M -
9 2
5
Mixer M-925
Instructions 56-8101-00 Edition AC