Formatting Output Data
Use the Data Formatting task pane of the DataMan Setup Tool to construct
a customized output string each time you read and verify a symbol. This
allows the verifier to return Code Quality results when connected to a third-
party application instead of the DataMan Setup Tool.
In addition to standard formatting, the DataMan Setup Tool supports the
use of Perl-Style Regular Expression features in Advanced mode.
22 DataMan 100 Verifier Quick Reference
Data Logging
Use the Data Logging task pane to generate a Code Quality Report for
each symbol that you read and verify. Be aware that data logging works
only when the reader is connected to the DataMan Setup Tool.
1.
Select
View->Advanced if the DataMan Setup Tool is not in
Advanced mode.
2.
Select the
Data Logging task pane under System Settings.
3.
Enable the
Report Storage option and use the Save in option, as
shown in the following example, to configure a directory location to
save each report along with an image of the verified symbol:
In addition, you can enable the Prefix filename with option to give
each Code Quality Report a set prefix.
DataMan 100 Verifier Quick Reference 23