Understanding Readings (20 Channels Shown For Clarity)
Pulse Artifact
Mild Power Line Artifact
Strong Power Line Artifact
12
EEG sensors are sensitive
enough to detect small
electrical signals on the
cortex. Occasionally, sensors
may lie directly atop a blood
vessel. When this occurs,
sensors may detect the
subject's pulse. Slightly move
the sensors to reduce pulse
artifacts.
Power lines operate between
50 and 60 Hz. Unsecured
power sources may leak
electromagnetic frequencies.
Because EEG records
frequencies from 1hz to about
40hz, power line noise may
negatively impact recordings
on all channels. This is
especially the case when
connection between sensor
and scalp is poor.
To avoid power line artifacts,
have subjects sitting at
least 3 feet from power
cords, plugged-in devices,
and lighting or overhead
fluorescent lights. The
Mobile-128 shields the
sensors from environmental
noise, but issues can still
occur with poor skin contact.
If power line artifacts appear
on all channels, lightly wet A1
Reference Drypad Ear Sensor.