Before unpacking your Model 8000, please note that although it is a durable industrial
scale, it is also a sensitive weighing instrument. Normal care should be taken when
handling and using the scale. Please observe the following precautions to insure years
of trouble free service from your Model 8000.
* DO NOT drop the scale platform or indicator.
* DO NOT drop objects onto the scale platform.
* DO NOT immerse the scale platform or indicator.
* DO NOT pick up the scale platform by the "spider"
Improper handling or abuse of the scale could be damaging and result in costly repairs
to your unit which may not be covered under warranty.
Carefully remove the scale platform, the indicator and the platter from the shipping
carton. Retain the shipping carton and all packing material in case reshipment is
required. Note that the indicator and platform are connected by a 5' cable; be careful
not to pull too hard on this cable.

LOCATING THE SCALE

Choose a location that is sturdy and free from vibration and air currents. Vibration
from heavy equipment and air currents from fans, heating ducts, etc. can cause the
scale display to fluctuate.
Locate the desired position for the scale indicator. Use the mounting bracket to
permanently mount the indicator to a wall, shelf or counter. Use the two holes
provided in the bracket to affix it to the desired location.
Adjust the indicator to its desired viewing angle and tighten down the knobs.
Place the scale platform in its desired location. Level the platform by adjusting the
four corner feet until the bubble in the bubble level, located beneath the top platter of
the platform, is within the center circle.
Test for a stable condition by trying to rock the platform backward and forward and
side to side. Adjust the feet for final levelness and stability.

ELECTRICAL CONNECTIONS:

The Model 8000 requires 115VAC, 50/60 Hz power (220VAC optional) for charging.
Be sure that the AC power is not too noisy - this can occur if large inductive loads,
like solenoids or motors are on the same power line. The 8000 has a filtered power
supply to reduce the effects of normal line noise, but it cannot limit severe noise.