4. Replace the top plate and sleeves.
Lamp Replacement Steps:
1. Allow the lamps to cool down before replacing a lamp.
2. Remove the top plate and lamp sleeves of the heating chamber to expose the IR lamps.
3. Find the malfunctioning IR lamp. Usually, the lamp will appear opaque/white when it burns
out.
4. Using clean cloves, hold the lamp by the ceramic white coating and remove from the lamp
holder.
5. Using a replacement lamp, hold the lamp by the ceramic white coating and align the lamp prongs
with the lamp holder. Once aligned insert the prongs into holder until flush and secured in place.
CB15S – Technical Specifications, Setup and Operation Manual
CBTI-MAN-001
Rev: 0
Page 4
Date Issued: 2019-11-13