In the spectrophotometer, light from the Halogen Lamp is focused on the entrance slit of
the Monochromator where the collimating mirror directs the beam onto the grating. The
grating disperses the light beam to produce the spectrum, a portion of which is focused
on the exit slit of the Monochromator by a collimating mirror. From here the beam is
passed to a Sample Compartment through one of the filters, which helps to eliminate
unwanted second order radiation from the diffraction grating. Upon leaving the Sample
Compartment, the beam is passed to the silicon photodiode Detector and causes the
Detector to produce an electrical signal that is shown on the Digital Display window.

Specifications

Table-1 lists the specification for both 1200 and 1201.
Table-1
Specifications
Wavelength Range
Spectral Bandpass
Wavelength Accuracy
Wavelength Repeatability
Stray Radiant Energy
Photometric Range
Photometric Accuracy
Power Requirements
Sample Holder
Dimensions
Net Weight
Model 1200
< 0.5%T at 340 and 400 nm
0~1999C (0~1999 Factor)
115/230 V ±10%, 60/50 Hz switchable
Single Test Tube/Cuvette
415W x 320D x 175H (mm) (16-1/2 x13 x 6-3/4 inches)
5
Model 1201
325~1000 nm
5 nm
± 2 nm
± 1 nm
0~125%T
0~2.0 Abs
± 1.0%T
4-poisiton Cuvette Holder
6.5 kg (15 lbs)