Canta V5 Kurzanleitung

Blättern Sie online oder laden Sie pdf Kurzanleitung für Medizinische Ausrüstung Canta V5 herunter. Canta V5 2 Seiten. Oxygen concentrator
Auch für Canta V5: Benutzerhandbuch und Wartung (3 seiten)

Canta V5 Kurzanleitung
Quick Guide - V5 / V8 OXYGEN CONCENTRATOR USER GUIDE AND MAINTENANCE - Filter Change
V5 / V8 OXYGEN CONCENTRATOR REAR VIEW
Air Inlet Filter access panel
Exhaust
(under base of unit)
AIR-INTAKE FILTER ACCESS PANEL
Dust Filter
DUST FILTER
The dust filter is located on
the rear of the Filter access
panel door.
AIR -INTAKE FILTER
The white air-intake filter is
Power cord
located in the air-intake
filter compartment via the
access panel.
The Air-intake filter needs to be replaced after every
1000 hours of use or annually (whichever comes first
for the first change). For subsequent filter changes the
device will have a label indicating total working hours
required before the next filter change. The device
needs returning to MESU for the filter changes.
During home use a replacement air-intake filter is
Air-intake
available from MESU.
filter
The Dust filter can be cleaned
with mild detergent and water,
rinse to remove all residues and
air dry.
The dust filter should be
inspected every 2 weeks and
replaced when it is visibly dusty.
HUMIDIFIER BOTTLE
Humidifier bottles are single patient use.
Bottles must be cleaned daily with soap and water.
Fill with distilled
water only.
Water must be kept
between Maximum
and minimum scale
line
Change the water
every 24hrs
Y TUBE CONNECTOR
Administer 15lpm from 2 x V8s