Cantel SB-110-HU1016 Handbuch - Seite 3

Blättern Sie online oder laden Sie pdf Handbuch für Medizinische Ausrüstung Cantel SB-110-HU1016 herunter. Cantel SB-110-HU1016 4 Seiten. Using endoscope flushing aid 24-hour multi-use tubing universal hookup with fujifilm™ 700 series endoscopes

Using SCOPE BUDDY
PLUS Endoscope Flushing Aid 24-Hour Multi-Use Tubing Universal Hookup (SB-110-HU1016) with FUJIFILM
3. Connect the red luer of the Universal Hookup (SB-110-HU1016) to the clear tube luer (suction channel) of the
WA-007 cleaning adapter. Connect one of the clear luers of the Universal Hookup (SB-110-HU1016) to the blue
tubing luer (air/water channel) of the WA-007 cleaning adapter
4. If the endoscope has a forward water jet, connect one of the clear luers to that channel. Clamp off any
remaining unused tubes. The Universal Hookup is now ready to be connected to the SCOPE BUDDY
Endoscope Flushing Aid pump tube.
If you have any questions or concerns regarding this product bulletin, please contact Customer
Care at 1-800-444-4729.
700 Series Endoscopes
PLUS
3