MAINTENANCE TROUBLESHOOTING
Pump won't turn on
1. Be sure you are pressing the button fully. There is a 'click' sound before the button is pressed enough to latch.
Pump motor won't turn
1. Check ow rate setting
2. Be sure that pump head is fully locked onto the faceplate bracket
3. Be sure that pump head is fully closed
4. If the foot pedal won't start the motor, use the prime button to determine whether the foot pedal or the internal switch is bad.
5. Try a foot pedal from a different pump. (Pumps with similar foot pedals include EGP-100 and Olympus OFP-1.)
6. Not all pump heads are interchangeable even though they may t. Be sure that the pump head is labeled "Endo Stratus
Irrigation Pump".
7. Remove the pump head and touch a magnet to the small screw on the face plate. The motor shaft should turn when the
prime button is pressed.
Water warmer not warming
1. The bottom of the warmer should be warm to the touch one minute after pressing the water warmer activation switch.
2. If the water warmer activation switch is blue when the feature is rst turned on (and cold), ensure that the water warmer
cable is plugged into the rear panel connector.
3. Note that warming a bottle from room temperature to body temperature will take over an hour. Water bottles should be
prewarmed.
Pump motor turns for some time when the foot pedal is pressed and then stops.
1. The foot pedal may have a leak. Submerge it in water and check whether bubbles come from the joint between the
top rubber and the bottom rigid part.
Insuf ator won't turn on
1. Be sure you are pressing the power button fully. There is a 'click' sound before the button is pressed enough to latch and
the button remains pressed in. .
Water warmer not warming
1. The bottom of the warmer should be warm to the touch one minute after pressing the water warmer activation switch.
2. If the water warmer activation switch is ashing blue, ensure that the water warmer cable is plugged into the rear panel
connector.
Pump Head Cartridge:
1. On/Off Button functions properly and actuates to turn pump on/off.
2. H2O button remains green to indicate proper functionality. It will turn blue if the heater is unplugged.
3. Explanations of the priming function. Button remains lit while priming only.
4. Flow rate knob should move freely without hesitation
Foot Pedal:
1. Attaches securely to the pump
2. Activates pump immediately upon depression.
3. Deactivates as soon as pressure is released from pedal.
Bottle Warmer:
1. Securely fastened to the base of pump
2. Connection is secure and heats properly.
3. Bottle ts securely in holder once wing nut is tightened.
CO2 tank drained while the insuf ator was not in use.
1. Check for leaks where the source hose connects to the rear panel of the EGA-501 and the source. Flare ttings can be
checked for leaks using a small amount of soapy water, taking care to prevent water from reaching the electrical
connections of the EGA-501 or any other equipment. If leak is detected, tighten the connection additionally by hand until
the leak stops. According to SAE J512, the recommended torque settings for the flare fitting are 10 to 11 N•m or 84 to 97 in•lbs.
2. Ensure that the ow rate setting is turned off when the Endo SmartCap is not connected to an endoscope.
Both of the internal valves are closed when the ow rate setting is turned off.
3. Ensure that the device is turned off when the device won't be used soon. Both internal valves are closed when the device
is turned off.
4. Turn off the source valve (at the wall or the tank) after the last procedure of the day. While every EGA-501 is tested to
ensure leak-free performance, a very small leak may exist in any high-pressure pneumatic system.