P0002W006 - MicroBio MB2-RSH Installation and Operation Supplement
Calibration!
!
The MB2-RSH can be calibrated in the same manner as any other MicroBio bioaerosol
sampler.!
!
Remove the remote sampling head from its normal location, disconnect the extension
cable from the control unit and then plug the remote sampling head directly into the control
unit as below:!
!
If extension cables over 10m in length are used, then it is recommended that the sampler
is calibrated with a cable of the installed length fitted to compensate for any voltage drops
on longer cables.!
!
For calibration specifics, please refer to document P0001W002 for calibration methodology
or if using the P0059 Validation Kit, refer to P0059W003.!
!
Issue 2!