Guidance and Manufacturer's Declaration - Immunity
The armband is intended for use in the electromagnetic environment specified
below. The customer or user of the armband should ensure that it is used in such an
environment.
IEC 60601
Immunity Test
Test Level
ESD
±6kV Contact
IEC 61000-4-2
±8kV Air
EFT
±2kV Mains
IEC 61000-4-4
±1kV I/Os
s
±1kV Differential
±2kV Common
Voltage Dips/
>95% Dip for
0.5 Cycles
Dropout
60% Dip for 5 Cycles
IEC 61000-4-11
30% Dip for 25
Cycles
>95% Dip for 5
Seconds
Power
3A/m
Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8

User Guide

Compliance
Level
Environment - Guidance
±6kV Contact
Floors should be wood, concrete, or ceramic
±8kV Air
tile. If floors are synthetic, the r/h should be
at least 30%.
N/A
Mains power quality should be that of a typi-
cal commercial or hospital environment.
N/A
N/A
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the armband
requires continued operation during power
mains interruptions, it is recommended that
armband be powered from an uninterrupt-
ible power supply or battery.
3A/m
Power frequency magnetic fields should
be that of a typical commercial or hospital
environment.
Electromagnetic
14