3.4 Connecting the encoder supply
Connect the encoder supply at terminals 19 and 20.
Do not use the encoder supply to supply non-earthed inductive or
capacitive loads.
The encoder power supply is not short-circuit proof.
Terminal
Voltage
19
0V
20
+24 VDC
+10 %/-50 %
3.5 Interface connect
The serial interface is capable of executing the following functions:
– Data printout
Interface parameter are as follows:
– Data transmission rate (baud rate),
– Parity bit,
– Number of stop bits,
The interface parameters can be set in the programming
(lines 43, 44, 45 and 46).
The following interfaces can be connected to the counter:
– RS232
Interface characteristics
RS232 Transmission with the following characterisics:
– Asymmetric
– Two leads
– Point-to-Point connection - 1 transmitter and 1 receiver
– Max. data transmission distance: 30 m
NE215
Max.
Max. permissible
residual ripple
current
Depending on load
80 mA