Doza MKS-15D Snegir Benutzerhandbuch - Seite 11
Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Messgeräte Doza MKS-15D Snegir herunter. Doza MKS-15D Snegir 17 Seiten. Dose-rate meter / radiometer
2.3.9.2 When the value of the set control threshold is reached, sound signal turns on. It is turned
off in the following cases:
-
if the value measured within 8 seconds doesn't exceed the set threshold;
-
at the shutter being opened/closed;
-
for 20 seconds - if any button is pressed; at that the button function is being operated; after
20 seconds the sound signal will resume.
2.3.10 Turning the display illumination on/off
To turn the display illumination on without changing the operation mode, press the
for a short time. The illumination will turned off automatically in 15 seconds. If within this period the
button is pressed again, the illumination will turn off.
2.3.11 Battery replacement
The dosimeter's power supply voltage is in the range 4 to 1.8 V. The dosimeter can operate
when the voltage drops down to 0.9 V, but take into account that during startup the dosimeter
consumes current of up to 50 mA, and thus fully discharged batteries can not provide the required
starting voltage. The foregoing also applies to the case when one battery is deeply discharged and the
other battery is fully operable.
In the upper part of the display window the residual capacity of batteries is indicated according
to the Table 2.2.
Table 2.2 – Residual capacity of batteries
Symbol
Capacity of
More than
batteries (%)
90 %
When
symbol appears on the display, batteries should be replaced. To replace batteries,
open the battery compartment and insert two alkaline batteries of AA-size in accordance with polarity.
The following batteries can be used in the dosimeter (see Table 2.3).
Table 2.3 – Types of batteries be used for power supply
Typical residual capacity voltage
Type of
battery
90 %
NiCd
2.68
NiMh
2.60
ZnCl
2.74
Alkaline
2.88
CZn
2.64
Please note:
-
potential and residual capacity depend on the type of batteries and their correct installation;
-
indicated residual capacity is for reference only, the dosimeter does not take any action if
the potential drops below acceptable limit;
-
residual capacity is calculated on the basis of measured potential of two tandem batteries,
and thus readings may not correspond to reality due to the deep discharge or malfunction of one of the
batteries.
90÷70 %
(for two batteries)
70 %
50 %
30 %
2.60
2.54
2.48
2.52
2.48
2.44
2.40
2.20
2.00
2.70
2.58
2.40
2.24
2.14
2.02
FVKM.412152.005RE
70÷50 %
50÷30 %
Typical
capacity (Ah)
R
=10
load
10 %
Ohm
2.30
1.3 ÷ 2.7
2.30
1.8 ÷ 2.4
1.90
0.5 ÷ 0.8
2.20
2.5 ÷ 3.0
1.90
0.3 ÷ 0.6
11
Less than
30÷10 %
10 %
specific attribute
for GP batteries
Marking
Marking
Green case
Marking, golden case
Grey case
button