5. Installation wiring

6.1.Instrument installation
The instrument shall be installed in a clean, dry, well
ventilated place without vibration。
1.There are 4 studs on the bottom plate for installation。
2 . Interface between measurement and water sample
collection 1/4" NPT Internal thread , You can also choose card
sleeve installation.。
Influent mode: Down in and up out, Keep the inflow
valve as small as possible,The best way is that water
does not produce bubbles。
6. 2. Connection mode
 L: 220AC
 N:220AC
 PE: Earthing
 V+,V-:DC 24V input
 A,B,P+,P-:Communication wiring with front end(A:
Green wire, B: Black wire, P+: Red wire, P-: Yellow wire)
 I+: 4-20mA+ output ( Four wire system current positive, No
need to provide 24V)
 I-:4-20mA- (Four wire system current negative, No need
to provide 24V)
 DO1,com1:Alarm1 relay action
 DO2,com2:Alarm2 relay action
 DO3,com3:unuse
Operating instructions for SD Series Online Color Meter