Follett REFNAT5-GD Installationshandbuch - Seite 6
Blättern Sie online oder laden Sie pdf Installationshandbuch für Kühlschrank Follett REFNAT5-GD herunter. Follett REFNAT5-GD 17 Seiten.
Cooling System Operation and Loading
Follett Edge Series refrigerators have a programmable set point range of 2 C to 8 C in 0.1 degree increments, but
are preset with a temperature set point of 5 C. The cooling system, as well as the display temperature, operate on
the cabinet air temperature.
Once the refrigerator has been powered on and maintaining your desired temperature, slowly introduce product
into the cabinet in small batches and allow the temperature to recover. This procedure should be performed until
the cabinet is full.
The cooling system is designed to pull to set point range within 30 minutes of initial power up.
The cooling system will recover in roughly 15 minutes of a 3 minute door opening. Avoid frequent door openings
and introduction of large amounts of thermal mass at one time.
Defrost
Follett Edge Series refrigerators have an automatic defrost cycle designed to keep the evaporator coil frost free.
Refrigerators do not require manual defrosting.
For every 18 hours of cumulative compressor run time, the controller will initiate a defrost cycle. During this time
the compressor, evaporator fan, and condenser fan power down while the defrost heater is energized to remove
accumulated frost from the coil. Upon termination of the defrost cycle, the cooling system will again become
operational. The defrost system is designed to keep the cabinet within 2 C to 8 C during defrost.
6
REFNAT5-SD, REFNAT5-GD