Probe is not reading correctly
• Check the membrane and replace with new membrane if it is clogged by
something.
• Clean the cathode and anode if they are tarnished.
• Check for significant concentration of H2S , SO2 , H2 , Neon and NO..because
some gases can interfere with DO readings.
Display disappears
Check whether the low battery indicator is displayed on before the display is
disappeared, if yes, replace with new batteries.
Error code
E 1: The probe is disconnected or damaged.
E 2: The value is underflow.
E 3: The value is overflow.
E 4: The original data that is related to this value error.
E17: 100% saturation Calibration error.
E21: The current temperature is out of +/- 10 C of the temperature when doing
100% saturation calibration. Perform 100% calibration again.
E31: A/D error.
E32: IIC memory error.