Amprobe 30XR-A Benutzerhandbuch - Seite 9
Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Multimeter Amprobe 30XR-A herunter. Amprobe 30XR-A 14 Seiten. Professional digital multimeter
Auch für Amprobe 30XR-A: Benutzerhandbuch (14 seiten)
RESISTANCE
(45 Hz – 500 Hz)
AC VOLTS
Ranges: 200 Ω, 2 kΩ, 20 kΩ, 200 kΩ,
Ranges: 200 m, 2 V, 20 V, 200 V, 600 V
2 MΩ, 20 MΩ
Accuracy: All ranges, ± (1.5 % rdg +
Accuracy:
4 dgts)
200 Ω to 200 kΩ ranges: ± (1.0 % rdg
Resolution: 100 μV in 200 mV range
+ 4 dgts)
Input impedance: 10 MΩ
2 MΩ ranges: ± (1.5 % rdg + 4 dgts)
Overload protection: 200 mV range:
20 MΩ range: ± (2.0 % rdg + 5 dgts)
600 V dc or 600 ac rms 15 seconds.
Resolution: 100 mΩ in 200 Ω range
Other ranges: 600 V dc or 600 V ac rms
DC CURRENT
Open circuit volts:
200 Ω range: 3.0 V dc
Ranges: 200 µA, 2 mA, 20 mA, 200 mA,
10 A
Other ranges: 0.3 V dc typical
Accuracy:
Overload protection: 600 V dc or
200 μA to 200 mA ranges: ± (1.5 % rdg
600 V ac rms
+ 1 dgt)
CONTINUITY
Audible indication: 75 Ω ± 25 Ω
10 A range: ± (2.0 % rdg + 3 dgts)
Resolution: 0.1 µA in 200 μA range
Response time: 100 ms
Burden voltage:
Overload protection: 600 V dc or
200 μA Range: 1 mV/ 1 μA
600 V ac rms
2 mA Range:
100 mV/ 1 mA
DIODE TEST
20 mA Range:
13 mV/ 1 mA
Test current: 1.0 mA (approximate)
200 mA:
4.6 mV / 1 mA
Accuracy: ± (1.5 % rdg + 3 dgts)
10 A:
40 mV / 1 A
Resolution: 0.001 V
Overload Protection:
Open circuit volts: 3.0 V dc typical
μA / mA input: F 0.25 A /600 V, Min.
Overload protection: 600 V dc or
I.R. 30 kA, (6.3x32 mm)
600 V ac rms
10 A input: F 10 A / 600 V, Min. I.R.
BATTERY TEST
100 kA, (10x38 mm) (
10 A for 4 minutes
Ranges: 1.5 V, 9 V
maximum followed by a 12 minute cooling
Accuracy: ± (3.5 % rdg + 2 dgts)
)
period
Resolution: 1 mV, 10 mV
(45 Hz – 500 Hz)
AC CURRENT
Load Test current:
Ranges: 200 μA, 2 mA, 20 mA,
1.5 V range: 150 mA typical
200 mA,10 A
9 V range: 5 mA typical
Accuracy:
Overload protection: 600 V dc or
200 μA to 200 mA ranges: ± (2.0 % rdg
600 V ac rms
+ 4 dgts)
NON-CONTACT VOLTAGE (NCV)
10 A range: ± (2.5 % rdg + 4 dgts)
AC Volts: 70 V to 600 V ac
Resolution: 0.1 μA in 200 μA range
Red LED and Audible Indicator
Burden voltage: See DC Current
REPLACEMENT PARTS
Overload Protection:
TL36 – Test Lead Set w/ Alligator clips
μA / mA input: F 0.25 A /600 V, Min.
FP375 - Fuse Pack 250 mA/600 V
I.R. 30 kA, (6.3x32 mm)
(4 each)
10 A input: F 10 A / 600 V, Min. I.R.
FP160 - Fuse Pack 10 A/600 V (2 each)
100 kA, (10x38 mm) (
10 A for 4 minutes
maximum followed by a 12 minute cooling
period
)
8