Overview
Safety note:
The appropriate national
safety regulations must be
observed when installing,
putting into operation,
and running this instrument.
Mounting and Wiring
Color
Signal
white
TTL pulse output
yellow
alarm output
red
power supply (+V)
black
common (0 V)
Alarm/Control Function
PSF210 is a programmable flow
controller designed for use with low-
viscose clear or translucent liquids,
transmitting IR light. The operating
principle is based on measurement
of liquid volume through light-weight
rotor, IR sensor and sophisticated
electronics for measurement and
control. PSF210 can be used as
a flow meter, flow transmitter with
pulse output, ON/OFF flow controller,
or alarm unit suitable for water
treatment and other applications.
Mounting
Install PSF210 in the pipe system
using flexible hoses with ½" union
nuts. Do not over-tighten!
Make sure the arrow marking
matches the flow direction!
Slowly fill the system to avoid
air damaging the PSF210 rotor.
Avoid uncontrolled air pressure
through the unit.
This may destroy the rotor!
Wiring
Connect PSF210 via the cable wires
as given on the left.
PSF210 provides a 'window' type
alarm/control output function.
Both AL and AH limits can be
programmed by the user.
Operating
Programming
For programming, use the hidden key
below the LED display.
Press and hold the key until ALo or AHi
appears on the display.
Release the key to setup AL/AH alarm limit.
Increase the blinking digit value by tapping
on the key and hold it to go to the next digit.
When all digits have been set, press and
hold the key to store the value in the device
memory and exit programming mode.
Important notes:
The fluid to be measured must be
compatible with the wetted parts!
Clean the medium-supply lines thoroughly
before use.
Use of 20
solid particles or medium contamination may
damage the flow sensor and/or influence
the measurement results!
Mind that the operating principle
is based on volume measurement,
i.e. air in water is considered medium.
Operating
When in operating mode, PSF210 indicates
current flow on the LED display within
specified range (see 'Specifications').
Alarm output stays off if the measured flow
is between AL and AH and goes on outside
alarm 'window' (see 'Alarm/Control Function').
Alarm LED lamp 'AL' lights when the alarm
output is on.
Display Messages
g+7 (over range) - display value over 999.
L_V (under range) - display value below -99.
OL (over load) - flow rate is over
specified range (see 'Specifications').
--- (initial check) - device initialization.
Sto (store) - settings are stored in memory.
m pre-filter is advisable because