TESTING
13. Next, remove the Pre-Filter (Image 14) in the
Air scrubber section and note the location of the
Circuit Breaker Button (Image 15) on the side of the
Retractable Power Cord on the Right side of the Air
Scrubber.
If the UVC bulbs are not on, turn on both UVC/
UVGI switches on the black power units (Image 16)
above the clear plastic shield. If they are still not
on, check that the door for the HEPA Filter is shut
on the outside of the cart. (Not shown)
Unshielded UVC light can cause eye damage, so there is an interlock switch on the
external HEPA door that will turn them off when changing the HEPA filter, even if the
operator forgets to turn them off. Go to the outside of the cart and open the HEPA
access door at the bottom of the rear panel by turning the latches and lowering it to be
sure the UVC lights and the Air Scrubber Fan turn off as it is opened.
WARNING!!
PAGE 10
(Image 14)
(Image 15)
(Image 16)