Dr. Honle dermalight 80 Betriebsanleitung - Seite 7

Blättern Sie online oder laden Sie pdf Betriebsanleitung für Medizinische Ausrüstung Dr. Honle dermalight 80 herunter. Dr. Honle dermalight 80 16 Seiten. Psoracomb

Dr. K. Hönle Medizintechnik GmbH
3.3.
E
LECTROMAGNETIC
dermalight
80 uses high frequency energy exclusively for its internal function. Therefore its
HF-emission is very low and it is unlikely that surrounding electronic devices are disturbed.
(HF-emissions acc. to CISPR 11, group 1)
dermalight
80 is determined for the use in an environment as described below. The user
should take care that the unit is used in such an environment.
dermalight
80 is determined for the use in all facilities, including residential areas and such
facilities connected directly to an official supply network, which also supplies buildings and
residential buildings. (HF-emissions acc. to CISPR 11, class B)
Magnetic fields at the supply frequency should correspond to the typical values for business
and hospital environments.
Portable and mobile radios should not be used at a lower distance to the dermalight
including cables than the recommended security distance acc. To the table below:
Nominal capacity of
the transmitter
0,01
0,1
1
10
100
C
OMPATIBILITY
Security distance depending on the transmitter frequency m
150 kHz - 80 MHz
0,03 m
0,11 m
0,35 m
1,11 m
3,50 m
dermalight
B
EDIENUNGSANLEITUNG
80 MHz - 800 MHz
0,03 m
0,11 m
0,35 m
1,11 m
3,50 m
6
®
80 uva
O
I
PERATING
NSTRUCTIONS
800 MHz - 2,5 GHz
0,07 m
0,22 m
0,70 m
2,21 m
7,00 m
80